- 电阻范围为1mΩ至10MΩ
- 额定功率:225 W / decade,最大250 W.或更高(联系IET定制更高功率单元)
- 准确度:0.5%或1%
- 额定电压:1000 V或更高
- 提供特殊和自定义配置
The High-Power Decade Resistance Substituter (HPRS) Series is a family of resistance decade boxes offering a broad choice of high-power, excellent-performance resistance sources.
The HPRS Series resistance decade boxes employ state-of-the-art precision resistors of various types for high accuracy, high stability, and low temperature and power coefficients of- resistance. High-power resistors within the unit are made available without sacrificing other electrical properties.
The standard models offer a choice of 3 to 8 decades. The panels are clearly labeled showing the step size and maximum voltage and current limitations for each decade.
With a resolution as low as 1 mΩ and a maximum resistance of 10 MΩ , the HPRS resistance decade box may be used as a general purpose substituter as well as a high power load for testing power supplies and for other high-power applications.
Applications include calibration of meters and instruments. HPRS resistance decade boxes are useful development tools wherever precise resistances with high-power handling capacity are required.
Accuracy:
Option F: ±(1% + 20 mΩ)
Option C: ±(0.5% + 20 mΩ)
Zero Resistance: <5 mΩ per decade
Maximum Voltage to Case: 1000 V peak Switches:
Continuous rotation 10 positions marked "0"-"9"
Resistance per step
|
Total decade resistance
|
Max current*
|
Max power per step* (W)
|
Temperature coefficient (ppm/°C)
|
Resistor Type
|
1 mΩ
|
0.009 Ω
|
8 A
|
0.036
|
±50
|
Resistance Wire
|
10 mΩ
|
0.09 Ω
|
6 A
|
0.36
|
±50
|
|
100 mΩ
|
0.9 Ω
|
6 A
|
3.6
|
±20
|
|
1 Ω
|
9 Ω
|
5 A
|
25
|
-20 to +80
|
Power Film
|
10 Ω
|
90 Ω
|
1.5 A
|
25
|
-20 to +80
|
|
100 Ω
|
900 Ω
|
0.5 A
|
25
|
-20 to +80
|
|
1 kΩ
|
9 kΩ
|
150 mA†
|
25†
|
-20 to +80
|
|
10 kΩ
|
90 kΩ
|
50 mA†
|
25†
|
-20 to +80
|
|
100 kΩ
|
900 kΩ
|
V limit†
|
V limit†
|
-20 to +80
|
Wirewound/metal film
|
1 MΩ
|
9 MΩ
|
V limit†
|
V limit†
|
±50
|
Wirewound/film
|
*Subject to maximum of 250 W max. per unit: 9 Steps per decade, multiply the decade value by the power per step. Multiple decades can be used to share power loads
|
|||||
†Subject to 1000 V (dc+ac) peak max
|